A finger joint rehabilitation device is composed of a finger holder, wrist support, face mask, base, electric motor, transmission mechanism, electronic circuit, and hand controller. The transmission mechanism consists of a rotating shaft, a worm gear seat, a worm gear, a worm wheel, a worm gear shaft, a large gear, an intermediate gear, and a fixed gear. This rehabilitation device can mimic the natural movement of fingers extending and bending during operation. It has the characteristics of labor-saving, low cost, and good effect in the rehabilitation treatment of finger joints. This rehabilitation device is suitable as a rehabilitation treatment device for patients with finger disabilities to passively burn their finger function, prevent finger joint rigidity and tendon adhesion after surgery.
Technical features:
1. Micro special motor drive, with reasonable mechanical structure design and material selection, high precision, smooth operation, low noise, and long service life.
2. Micro special computer digital control design, using imported integrated components, equipped with a large capacity imported CPU central processing unit, intelligent software programming, and functions such as over torque overload protection.
3. The large screen LCD backlight display provides a clear view of all parameters, making it convenient for medical clinical research.
4. The wide range of joint mobility and adaptability increase, fully considering the needs of medical treatment.
5. Adopting optimized structural design, the distribution of joint angle and velocity changes is reasonable.
6. The safety handguard is set up to ensure that the palm will not be injured due to machine malfunctions.
Technical parameters:
1. The last set data appears first when the rehabilitation device is turned on
2. The adjustable length range of the rehabilitation device activity bracket is not less than 50mm.
3. The angle (α) between the fixed bracket and the movable bracket of the rehabilitation device changes within a range of not less than 100 ° in terms of movement
4. The rated load of the rehabilitation device is 50N, and it should be able to work smoothly without jamming under the rated load, and there should be no abnormal impact sound during reciprocating operation.
5. The working noise of the rehabilitation device should not exceed 65dB.
6. The range of angular velocity adjustment for the angle (α) between the fixed and movable supports of the rehabilitation device is: the lowest speed is not greater than 0.5 °/s, the highest speed is not less than 1.5 °/s, and it can be adjusted in stages.
7. The rehabilitation device can work continuously for more than 4 hours under a load of 50N.
8. The rehabilitation device is equipped with a manual controller, allowing patients to control the device to pause or perform a reset movement (alpha to 180 ° movement) on their own.
9. The input power of the rehabilitation device is 70VA.
10. Rated working voltage: 220VAC ± 22VAC